MCA166-USB
Command name CMD_SET_REPEAT
Description The command sets the number(rep: 0 ... 65535) of sweeps for repetitive
measurement.
Command syntax
Format integer integer integer long integer
Parameter preamble command rep 0 end flag
Byte String (HEX) A5 5A 4A 00 rep
l
rep
h
00 00 00 00 B9 9B
Used setup value None
Affected setup value Repeat
MCA state STATE_READY
Command name CMD_SET_MCS_CHANNEL
Description The command set the number of channels(ch: 128 ... 4096) for MCS mode
Command syntax
Format integer integer integer long integer
Parameter preamble command ch 0 end flag
Byte String (HEX) A5 5A 63 00 ch
l
ch
h
00 00 00 00 B9 9B
Used setup value None
Affected setup value mcs_channels
MCA state STATE_READY
Command name CMD_SET_TIME_PER_CHANNEL
Description The command sets the dwell time(tpc: 1 ... 65535) per channel for the MCS
mode
40
.
Command syntax
Format integer integer integer long integer
Parameter preamble command tpc 0 end flag
Byte String (HEX) A5 5A 4B 00 tpc
l
tpc
h
00 00 00 00 B9 9B
Used setup value None
Affected setup value time_per_channel
MCA state STATE_READY
40 It is the real time per channel multiplied with 10